Temperature
Range:
6°C above ambient to +260°C.
Uniformity:
±3.5% of setpoint, nominal. Actual performance may vary depending
on load size and placement, chamber size, ambient temperature,
atmosphere and environmental conditions.
Controller:
Single setpoint digital microprocessor control, front-mounted.
Simultaneous LED display of setpoint and actual temperature.
Overtemperature:
Independent overtemperature safety system factory set with
reset pushbutton.
Heating
Element:
Low watt design for extended life performance.
Vacuum Capability:
1 x 10-2 torr, (10 microns)
Vacuum
System:
1"
(2.54 cm) manifold, chamber rear, exterior, connects to pump
or in-house vacuum source. Includes inert gas port, vacuum
port, and vent port.
Atmosphere:
For use with non-flammable, non-corrosive inert
gases only. Vacuum gauge, vacuum, gas and vent valves located
on front.
Interior:
Corrosion resistant stainless steel with aluminum
shelves.
|
Cabinet:
Heavy
gauge steel construction, dual wall, with fiberglass insulation
for minimum heat loss in walls and door. Painted exterior
or deluxe stainless steel finish.
Gasket:
High
temperature silicone specially molded for positive seal.
Double
Door:
Solid
outer door over inner window door is standard on stainless
steel exterior models.
Positive
Door Seal:
Magnetic
latch. V01824 Series units include mechanical latches.
Viewing
Window:
Standard
1/2" thick on all models; permits observation of work in process.
V01824 is 3/4" (1.9 cm) thick.
Shelves:
V0914 and V01218 Series shelves are removable;
V01824 Series shelves are fixed.
Accessories:
Pumps,
traps, hoses, clamps and fittings may be ordered separately.
Electrical:
See
specifications for details. Power cord included on all models.
UL
Listing:
Models
V0914A, V0914SA, V01218A, V01218SA.
|